Closed Solicitation · DEPARTMENT OF HEALTH AND HUMAN SERVICES

    RETROFIT GLATT GCG 70 DRY MIXER FOR SIGNAL EXCHANGE TO THIRD PARTY EQUIPMENT

    Sol. FDA-RFQ-25Q1283892Combined Synopsis/SolicitationROCKVILLE, MD
    Closed
    STATUS
    Closed
    closed Aug 1, 2025
    POSTED
    Jul 18, 2025
    Publication date
    NAICS CODE
    334516
    Primary industry classification
    PSC CODE
    6640
    Product & service classification

    AI Summary

    The FDA seeks a contractor to retrofit a Glatt GCG 70 dry mixer for enhanced communication with supervisory control systems. This project includes both hardware and site services, ensuring integration into continuous manufacturing operations. The contract is firm fixed price, with a performance period of 180 days post-award, and requires comprehensive installation and commissioning support.

    Contract details

    Solicitation No.
    FDA-RFQ-25Q1283892
    Notice Type
    Combined Synopsis/Solicitation
    Posted Date
    July 18, 2025
    Response Deadline
    August 1, 2025
    NAICS Code
    334516AI guide
    PSC / Class Code
    6640
    Primary Contact
    Warren Jackson
    State
    MD
    ZIP Code
    20857
    AI Product/Service
    both

    Description

    Background

    The Office of Pharmaceutical Quality Research (OPQR) supports the modernization of pharmaceutical manufacturing by conducting regulatory science research on emerging technologies. Continuous manufacturing (CM) of oral solids has been recognized as an emerging technology by the FDA. Continuous process monitoring and manufacturing process control are critical for establishment of a robust CM process that can consistently produce product of desired quality, while identifying and mitigating manufacturing risks.

    Currently, OPQR possesses a Glatt GCG 70 continuous dry mixer. However, the GCG 70 lacks the capability to communicates with other equipment or systems. To integrate the GCG 70 into a CM line, a retrofit is required to enable it to interface with a supervisory control system. This upgrade will enhance the functionality of the GCG 70.

    This package includes both hardware and site services for retrofitting the GCG 70 dry mixer to facilitate signal exchange with other equipment. The retrofitted GCG 70 should be adaptable to CM operation which includes options for the seamless integration of upstream and downstream unit operations. It will also enable synchronized control of process parameters of other unit operations through automatic modulation.

    This retrofit is essential to support the New Drug Application (NDA) and Abbreviated New Drug Application (ANDA) review processes for CM applications.

    Scope

    To retrofit the GCG 70 Dry Mixer to communicate with the other equipment of systems.

    Requirements

          The GCG 70 retrofit shall meet the following salient characteristics:

    • Shall enable electronic communication with SIMATIC PCS 7 distributed control system for integration with other equipment.
    • Shall enable setting blender speed remotely, including start/stop, as well as feedback indicating that the blender is running and the actual speed.
    • Shall not restrict or inhibit any existing operational capabilities of the GCG 70, including local control capability.
    • Shall include one (1) year parts and labor with warranty from date of shipment.

    Delivery, Installation, and On-site Commissioning

               The vendor will cover all costs associated with the following:

    • Hardware packaging, shipping, and delivery.
    • On-site hardware installation, commissioning, and qualification.
    • On-site assistance with supervisory control system commissioning.
    • Updated qualification documents, instruction manuals, drawings/diagrams, and other related equipment documentation for the GCG 70

      1. INSPECTION/ACCEPTANCE CRITERIA

    Inspection and acceptance shall be in accordance with the base clause 52.212-4(a). Contracting officer’s representative (COR) is responsible for the inspection and acceptance of the deliverables for this requirement.

     CONTRACT TYPE

    This is firm fixed price.

     PERIOD OF PERFORMANCE

     The period of performance (POP) is for 180 days after award.

     PLACE OF DELIVERY AND PERFORMANCE

    U.S. Food and Drug Administration

    5901 Ammendale Road

    Suite G

    Beltsville, MD 20705

    Attn: POC to be identified at time of the award.

    Key dates

    1. July 18, 2025Posted Date
    2. August 1, 2025Proposals / Responses Due

    AI search tags

    Frequently asked questions

    RETROFIT GLATT GCG 70 DRY MIXER FOR SIGNAL EXCHANGE TO THIRD PARTY EQUIPMENT is a federal acquisition solicitation issued by DEPARTMENT OF HEALTH AND HUMAN SERVICES. Review the full description, attachments, and submission requirements on SamSearch before the response deadline.

    SamSearch Platform

    Stop searching. Start winning.

    AI-powered intelligence for the right opportunities, the right leads, and the right time.