Active Solicitation · DEPARTMENT OF HEALTH AND HUMAN SERVICES
AI Summary
The National Institutes of Health is seeking to procure two Triple Multitron Standard Incubation Shakers with Top Cooling. These units must meet specific requirements for temperature control, capacity, and design to fit laboratory space. Bidders should ensure compliance with the detailed specifications outlined in the solicitation.
NIDDK owns two Triple Multitron Standard Incubation Shakers with Top Cooling. Each Triple Multitron consists of three stacked incubation shakers within a small footprint. Each of the three incubation shakers within a Triple Multitron is capable of independent temperature and speed control. These features maximize the total amount of cultures that can be grown at the same time, as well as, maximizing the possible growth conditions. Currently, none of the incubation shakers within the two Multitrons cool properly; and they do not reliably maintain a constant temperature. This procurement is to buy two new Triple Multitron Standard Incubation Shakers with Top Cooling.
The following product features/characteristics are required for this requirement:
Each triple incubation shaker unit must consist of three stacked incubation shakers in a small footprint to fit within the available space in the laboratory (not wider than 43 inches and not deeper than 35 inches)
Each incubation shaker per triple stacked unit must be capable of independently reaching a temperature that is 15 degrees C below ambient temperature to 65 degrees C to accommodate a large range of temperatures to grow cultures.
Each incubation shaker per triple stacked unit must have a capacity of 6 x 5L Erlenmeyer flasks (total of 18 flasks per triple-stacked unit) to maximize the volume of cultures grown within the small footprint.
The temperature and speed of each incubation shaker per triple stacked unit must be able to be set independently of the others to maximize the variety of growth conditions within the small footprint.
Each incubation shaker must have a single bearing eccentric drive to move the main shaft for quiet operation and very little vibration.
Each incubation shaker must have the entire shaking mechanism enclosed to ensure temperature uniformity.
Each incubation shaker must have four blowers to eliminate dead spots within the cabinet to maintain temperature uniformity.
The height of the triple stacked unit must allow it to fit through standard laboratory doors.