    ANNUAL PREVENTIVE MAINTENANCE AND CERTIFICATION FOR BIOLOGICAL SAFETY CABINETS (CLASS II, TYPE A2), CHEMICAL FUME HOOD, AND PCR CABINETS/CLEAN BENCHES

    Sol. N6311626Q0516Sources SoughtSAN DIEGO, CA
    Closed
    STATUS
    Closed
    closed Feb 17, 2026
    POSTED
    Feb 10, 2026
    Publication date
    NAICS CODE
    541380
    Primary industry classification
    PSC CODE
    H266
    Product & service classification

    AI Summary

    The Department of Defense is seeking sources for annual preventive maintenance and certification of laboratory equipment, including Biological Safety Cabinets, a Chemical Fume Hood, and PCR Cabinets. Interested vendors should provide capabilities information and reference N6311626Q0516. The work will take place in San Diego, CA.

    Contract details

    Solicitation No.
    N6311626Q0516
    Notice Type
    Sources Sought
    Posted Date
    February 10, 2026
    Response Deadline
    February 17, 2026
    NAICS Code
    541380AI guide
    PSC / Class Code
    H266
    Primary Contact
    David L. Hill
    State
    CA
    ZIP Code
    92106
    AI Product/Service
    service

    Description

    The Government is seeking responsible sources capable of performing annual preventive maintenance and certification for Biological Safety Cabinets (Class II, Type A2), a Chemical Fume Hood, and PCR Cabinets/Clean Benches. See below scope of work.

    SCOPE

    The vendor shall provide all necessary labor, tools, materials, and equipment to perform annual preventive maintenance (PM) and certification for a total of twenty-two (22) pieces of laboratory equipment located at the Naval Health Research Center. The equipment is as follows:

    • Biological Safety Cabinets (Class II, Type A2) – Qty: 18
    • Chemical Fume Hood – Qty: 1
    • PCR Cabinets/Clean Benches – Qty: 3

    The contractor's performance shall include, but not be limited to, the following services for each piece of equipment as applicable:

    Biological Safety Cabinets (Class II, Type A2):

    • Testing and certification in accordance with NSF/ANSI 49 standards.
    • Services must include downflow velocity profiling, inflow velocity measurement, HEPA filter integrity testing (leak scan), smoke pattern testing for airflow visualization, and site installation assessment.
    • Alarm function verification (e.g., airflow and window position alarms).
    • Decontamination of the work area before and after service.

    Chemical Fume Hood:

    • Testing and certification in accordance with OSHA 1910.1450 and CAL OSHA Title 8 5154.1.
    • Services must include face velocity measurement, smoke visualization testing to verify containment, and alarm function verification.

    PCR Cabinets / Clean Benches:

    • Testing and certification to ensure ISO 5 (Class 100) air quality conditions are met.
    • Services must include HEPA filter integrity testing and non-viable particle counting.
    • UV light function verification.

    Upon completion of service, the vendor shall provide the following:

    • Certification Reports: A detailed service report must be provided for each individual piece of equipment. Each report must clearly document the tests performed, the standards used, the results of each test (pass/fail), and any adjustments made.
    • Certification/Testing Labels: A physical label must be provided and affixed to each individual piece of equipment. The label must indicate date of testing/certification and the next scheduled certification date.

    The location of this requirement is San Diego, CA 92152.

    Interested sources may submit capabilities information by email to the point of contact listed below.  Reference N6311626Q0516 on submissions and include Company Name, Full Address, CAGE Code and Point of Contact information including Name, Phone Number and E-Mail address.

    Key dates

    1. February 10, 2026Posted Date
    2. February 17, 2026Proposals / Responses Due

    AI search tags

    Frequently asked questions

    ANNUAL PREVENTIVE MAINTENANCE AND CERTIFICATION FOR BIOLOGICAL SAFETY CABINETS (CLASS II, TYPE A2), CHEMICAL FUME HOOD, AND PCR CABINETS/CLEAN BENCHES is a federal acquisition solicitation issued by DEPT OF DEFENSE. Review the full description, attachments, and submission requirements on SamSearch before the response deadline.

    SamSearch Platform

    Stop searching. Start winning.

    AI-powered intelligence for the right opportunities, the right leads, and the right time.