Closed Solicitation · DEPT OF DEFENSE

    CARDIAC OUTPUT MONITORING SYSTEM

    Sol. HT9426-26-RFI-COMSSpecial Notice
    Closed
    STATUS
    Closed
    closed Jul 27, 2026
    POSTED
    Jul 22, 2026
    Publication date
    NAICS CODE
    339113
    Primary industry classification
    PSC CODE
    6515
    Product & service classification

    AI Summary

    The Defense Health Agency is conducting market research for a cardiac output monitoring system. This RFI seeks information from qualified contractors capable of providing a device that meets specific technical and operational requirements. Interested parties must submit their responses by 5:00 PM Eastern Time on July 27, 2026, via email to the designated contacts.

    Contract details

    Solicitation No.
    HT9426-26-RFI-COMS
    Notice Type
    Special Notice
    Posted Date
    July 22, 2026
    Response Deadline
    July 27, 2026
    NAICS Code
    339113AI guide
    PSC / Class Code
    6515
    Primary Contact
    Tammy McRae
    AI Product/Service
    product

    Description

    This is a REQUEST FOR INFORMATION (RFI) ANNOUNCEMENT ONLY. In accordance with Revolutionary Federal Acquisition Regulation Overhaul 15.101(c), the Government does not intend to award a contract based on this RFI or otherwise pay for information requested. The information will be treated as information only and not as a proposal. This is not a solicitation announcement nor a request for quotes and does not obligate the Government to award a contract. Requests for a solicitation will not receive a response. Responses to this RFI must be in writing.

    The purpose of this RFI announcement is to conduct market research to make appropriate acquisition decisions and to gain knowledge of potential qualified offerors interested and capable of providing the “Cardiac Output Monitoring System” described below. Documentation of technical expertise must be presented in sufficient detail for the Government to determine that your company possesses the necessary functional area of expertise and experience to fulfill the requirement.

    The Defense Health Agency Contracting Activity, 808 Schreider Street, Fort Detrick, MD 21702, is seeking to understand the capabilities of potential qualified Contractors who can provide the supplies and/or services for the following:

    Minimum Essential Characteristics: Cardiac Monitoring Platform

    The United States Army Institute of Surgical Research (USAISR) seeks a device that can monitor cardiac output (CO) and derive hemodynamic parameters in a hospital setting. The USAISR requires a CO monitoring device that utilizes a Swan-Ganz catheter to monitor hemodynamic parameters. The device’s proprietary coded output must be compatible with USAISR’s unique internal data logging systems. This is required to maintain historic and current protocol data continuity and workflow. Inability to meet these requirements will inhibit comparisons of current and future studies to USAISR historical data resulting in loss of data, and efforts.

    The CO Monitoring Platform shall have the following specifications:

    1. Ability to measure the following: mixed venous oxygen saturation (SvO2), CO, stroke volume, systemic vascular resistance, right ventricular ejection fraction, right ventricular end diastolic volume, pulmonary artery pressure, and pulmonary artery occlusion pressure.
    2. Utilization of the Edwards Lifesciences 7.5 FR Swan-Ganz CCOmbo V set or a Swan-Ganz of equal functionality and performance that provides continuous monitoring of SvO2 and measures continuous calculation of cardiac output by using thermodilution principles.
    3. Equipment dimensions equal to or smaller than 342 x 388 x 208 mm (height x width x depth)
    4. Equipment shall weigh less than 22 lbs
    5. Equipment ingress protection of at least IPX1
    6. Display size of at least 15 inches.
    7. At least two (2) COM1 serial port connectors, RS-232
    8. Operational environmental specifications:
      1. Temperature range: 0 to 37℃
      2. Relative humidity: 10-90% non-condensing
      3. Altitude: 0 to 3,000 meters
    9. Transportation environmental specifications:
      1. Temperature range: -18 to 45℃
      2. Relative humidity: 20 to 90% non-condensing
      3. Altitude: Maximum 6,096 m for up to 8 hours
    10. Monitor battery must have these technical characteristics:
      1. Output voltage (nominal) 14.4 V
      2. Maximum discharge current 4.096 A (8.5 A at 25℃)
      3. Cells 8 x Li-Ion (lithium ion)
    11. Monitoring platform technical specifications:
      1. Input/Output
        1. Projective capacitive touch screen
        2. At least two (2) RS-232 serial ports: Maximum data rate = 57.6 kilo baud
        3. At least three (3) USB ports: 2 x USB 3.0; 1 x USB 2.0
        4. At least one (1) RJ-45 ethernet port
        5. At least one (1) HDMI port
      2.  Heart Rate average display
        1. Response time: 40 seconds for step increase from 80 to 120 beats per minute (BPM), 29 seconds for step decrease from 80 to 40 BPM.
        2. CO monitoring on: Averaging time: Time between CO measurements (3 to 21 minutes); Update rate: Approximately 1 minute; Response time: 175 seconds for step increase from 80 to 120 BPM, 176 seconds for step decrease from 80 to 40 BPM.
      3. Electrical
        1. Rated supply voltage: 100 to 240 Vac; 50/60 Hz
        2. Rated input: 1.5 to 2.0 Amps
        3. Fuses: T 2.5AH, 250V; High breaking capacity; Ceramic
      4. Alarm
        1. Sound pressure level: 45 to 85 dB(A)
    12. Utilization of a Swan-Ganz patient cable with the following parameter measurement specifications:
      1. Continuous Cardiac Output
        1. Range: 1 – 20 Liter/minute
        2. Reproducibility: ±6% or 0.1 Liter/minute, whichever is greater
        3. Maximum thermal filament surface temperature 48℃
      2. Intermittent (bolus) Cardiac Output
        1. Range: 1 – 20 Liter/minute
        2. Reproducibility: ±3% or 0.1 Liter/minute, whichever is greater
      3. Blood temperature
        1. Range: 15 to 45℃
        2. Accuracy: ±3℃
    13. Roll Stand:
      1. Compatible rolling stand with bracket and plates for cardiac output monitoring system
        1. Base Dimensions: 12-15 inches wide x 10-12 inches deep to allow
        2. Height: Non-specific but need to be adjustable
    14. Warranty and Product Support:
      1. 12-month standard warranty
      2. 24-month product support and protection plan

    The North American Classification System code for this acquisition is 339113 –

     Surgical Appliance and Supplies Manufacturing (Size Standard: 800 Employees).

    Notice to potential offerors: All offerors who provide goods or services to the United States Federal Government must be registered in the System for Award Management (SAM) at www.sam.gov and complete online representations and certifications in SAM.

    All interested parties shall submit information by email to both Tammy McRae and Justin Pickett at tammy.n.mcrae.civ@health.mil and justin.h.pickett.civ@health.mil, no later than 5:00 PM Eastern Time on 27 July 2026.

    Key dates

    1. July 22, 2026Posted Date
    2. July 27, 2026Proposals / Responses Due