Closed Solicitation · DEPARTMENT OF ENERGY

    LABORATORY STEAM STERILIZER

    DEPARTMENT OF ENERGY
    Sol. 6-B185-Q-00354-00SolicitationLemont, IL
    Closed
    STATUS
    Closed
    closed Jul 22, 2026
    POSTED
    Jul 17, 2026
    Publication date
    NAICS CODE
    334516
    Primary industry classification
    PSC CODE
    6640
    Product & service classification

    AI Summary

    The Department of Energy is soliciting proposals for a laboratory steam sterilizer, including on-site installation and removal of the old unit. Proposals are due by 5:00 PM CST on July 22, 2026, and must include a fully executed ANL-70 Request for Quotation. This opportunity will be evaluated based on the lowest price, technically acceptable methodology.

    Contract details

    Solicitation No.
    6-B185-Q-00354-00
    Notice Type
    Solicitation
    Posted Date
    July 17, 2026
    Response Deadline
    July 22, 2026
    NAICS Code
    334516AI guide
    PSC / Class Code
    6640
    Primary Contact
    Laura Sanchez
    City
    Lemont
    State
    IL
    ZIP Code
    60439
    AI Product/Service
    product

    Description

    *THIS RFQ WILL BE EVALUATED BASED ON LOWEST PRICE, TECHNICALLY ACCEPTABLE METHODOLOGY*

    Looking to source a laboratory steam sterilizer with on-site installation and removal of old unit.

    Please review all applicable documentation and provide a proposal response no later than 5:00 PM CST July 22, 2026.

    Within your proposal, please include a fully executed copy of the ANL-70 Request for Quotation.

    Any questions can be directed to the Contracting Officer.

    Thank you in advance for your participation.

    Laboratory Steam Sterilizer Specifications

    Internal Volume:

    • 9.7 cu. ft. (275 L)

    Door Configurations:

    • Single Door, Pass-through

    Design Operating Temperature:

    • Gravity/Vacuum Cycles: 230°–275°F (110°–135°C)
    • Liquid Cycles: 219°–275°F (104°–135°C)
    • Optional Low Temp Cycle: 169°–212°F (76°–100°C)

    Interior Dimensions (w × h × d):

    • 21" × 21" × 38" (532 × 532 × 965 mm)

    Exterior Dimensions (w × h × d):

    • Single Door: 30" × 74" × 68.125" (762 × 1880 × 1222 mm)

    Door Operation:

    • manual (standard)

    Steam Source:

    • facility/house steam (standard)

    Alr Removal:

    • ejector (standard)

    Recesses and Barrlers:

    • freestanding (cabinet enclosure panels required)

    Key dates

    1. July 17, 2026Posted Date
    2. July 22, 2026Proposals / Responses Due

    AI search tags

    Frequently asked questions

    LABORATORY STEAM STERILIZER is a federal acquisition solicitation issued by DEPARTMENT OF ENERGY. Review the full description, attachments, and submission requirements on SamSearch before the response deadline.

    SamSearch Platform

    Stop searching. Start winning.

    AI-powered intelligence for the right opportunities, the right leads, and the right time.