Active Solicitation · DEPT OF DEFENSE

    MICROWAVE THERAPY SYSTEM

    Sol. W81K0426QA059Sources SoughtJBSA FT SAM HOUSTON, TX
    Open · 8d remaining
    DAYS TO CLOSE
    8
    closes May 21, 2026
    POSTED
    May 12, 2026
    Publication date
    NAICS CODE
    334510
    Primary industry classification
    PSC CODE
    6515
    Product & service classification

    AI Summary

    The Department of Defense is seeking information on a microwave therapy system for soft tissue ablation. Interested parties should provide system descriptions and specifications by May 21, 2026. This opportunity is a Sources Sought notice, and responses will not be compensated. Registration in SAM is required for participation.

    Contract details

    Solicitation No.
    W81K0426QA059
    Notice Type
    Sources Sought
    Posted Date
    May 12, 2026
    Response Deadline
    May 21, 2026
    NAICS Code
    334510AI guide
    PSC / Class Code
    6515
    Issuing Office
    W40M USA HCA
    State
    TX
    ZIP Code
    78234-5074
    AI Product/Service
    product

    Description

    This is a Request for Information (RFI) – Sought Notice Only. This notice is issued solely for information and planning purposes only.
     

    1. The US Government will not pay for any information or administrative costs incurred in response to this notice. Submittals will not be returned to the responder; not responding to this notice does not preclude participation in any further RFQ; IFB or RFP if issued.
     

    2. The associated North American Industry Classification System (NAICS) is 334510. The business size standard is 750 Employees. All contractors doing business with the Department of Defense shall be registered and approved through the System for Award Management (SAM). This is the Official U.S. Government system that consolidates the capabilities of CCR/FedReg, and EPLS. There is no fee to register for this site. Entities may register at no cost directly from this page. The proposed performance start date will be on or about 1 July 2026. The Solicitation number is W81K0426QA043.
     

    3. Note – Prior to award your company shall reflect the NAICS Code 334510 identified on your SAM. If your company is interested in competing on this project, and have services, please submit the following documentation no later than  21 May 2026, 10:00 AM EST, to Cynthia A. Richardson-Rhone and Jason Holck at cynthia.a.richardson-rhone.civ@health.mil


    Your email response shall provide the following:

    Can you provide the following:

    General System Description

    The Intelli Blate Microwave Ablation System is a software-controlled microwave therapy delivery platform intended for the coagulation (ablation) of soft tissues. The system utilizes microwave energy to produce controlled, spherical ablation zones and features integrated cooling, real-time temperature monitoring, and advanced probe placement functions.

    Generator (Console) Specifications

    • Operating Frequency  2.45 GHz ± 50 MHz
    • Ablation Channels       Must support at least two (2) independent ablation channels, allowing for single-probe, dual-independent, or dual-linked probe operations
    • Output Power  Single Channel: 0 to 90 Watts. Dual Channel: 0 to 80 Watts per channel (up to 160 Watts linked)
    • Power Adjustments     Output power must be adjustable in 5-Watt increments.
    • Safety Shut off            Must feature an automatic system shutdown if probe temperatures or reflected power exceed established safety limits
    • Dimensions & Weight Console alone must weigh no more than 40 lbs. (approx. 38 lbs.) for portability and mount securely to a customized mobile cart

    User Interface & Display

    • Monitor            15.6-inch foldable, high-resolution color touchscreen
    • Ergonomics     Screen must be capable of swiveling 180 degrees (CW/CCW) to allow visibility from any operating position. Must be fully operable with gloved hands or through a sterile drape
    • Visual Feedback         Interface must provide real-time tracking of time, tip power, reflected energy (power), and multi-point temperature data

    Integrated Cooling System

    • Pump Mechanism       Must feature two (2) fully integrated peristaltic pumps built directly into the console (one for each ablation channel)
    • Coolant Compatibility Must be compatible with standard, user-supplied sterile water or 0.9% saline (1000 mL bags or greater)
    • Cassette Loading        Must utilize a drop-in, integrated fluid cassette system to streamline coolant delivery and prevent setup errors

    Probe Assembly Characteristics (Ximitry Probes)

    • Antenna Profile           13.5 Gauge (2.3 mm diameter) fiberglass shaft with a ceramic trocar tip
    • Available Lengths       Must be available in at least three lengths: 15 cm, 20 cm, and 27 cm
    • Integrated Temperature Sensing        Probe must contain at least three (3) integrated thermocouples along the shaft and two (2) internal temperature sensors within the handle for continuous, real-time thermal feedback
    • Cable Management    Must utilize a single-cable bundle that integrates both the RF power cable and the coolant tubing (supply and return lines) into one cord
    • Visual Indicators         Probe handles must include integrated LED indicators to visually signal when microwave energy is actively being delivered

    Advanced Clinical Workflow Features

    • Needle Lock    Must include a feature that delivers a short, calculated burst of microwave energy to secure the radiating section of the probe to the tissue, mitigating accidental probe migration during setup or intra-procedural imaging
    • Tract Ablation  Must include a specific track ablation/cauterization mode to safely ablate the tissue along the insertion tract (4-10 mm) during probe retraction to prevent bleeding and tumor seeding.
    • External Temperature Monitoring       System must support the connection of up to two (2) remote/external Temperature Sensor Probes to monitor collateral tissue temperatures independently from the ablation probes
    • System Integration      Must be capable of integrating with advanced 3D planning and laser guidance software (e.g., my Ablation Guide, my Needle Companion) and support laser alignment discs on the probe handles.

    Please ensure to address the below:


    Name of Company
    Point of contact
    Phone number
    Email address
    Identify Your Small Business Socio-Economic Group
    DUNS number and CAGE Code


    4. This requirement No telephonic inquiries will be responded to. A solicitation will be posted within 2-3 weeks www.Sam.gov.

                   

    Key dates

    1. May 12, 2026Posted Date
    2. May 21, 2026Proposals / Responses Due

    AI search tags

    Frequently asked questions

    MICROWAVE THERAPY SYSTEM is a federal acquisition solicitation issued by DEPT OF DEFENSE. Review the full description, attachments, and submission requirements on SamSearch before the response deadline.

    SamSearch Platform

    Stop searching. Start winning.

    AI-powered intelligence for the right opportunities, the right leads, and the right time.