    AI Summary

    The Department of Defense is seeking vendors for a multi-therapy infusion pump designed for military and aeromedical use. The device must be compact, lightweight, and durable, with specific performance and environmental requirements. Interested vendors should submit a capability statement demonstrating compliance with the outlined specifications by the specified deadline.

    Contract details

    Solicitation No.
    FA4417MTPUMP
    Notice Type
    Sources Sought
    Set-Aside
    Total Small Business Set-Aside (FAR 19.5)
    Posted Date
    April 8, 2026
    Response Deadline
    April 14, 2026
    NAICS Code
    339112AI guide
    PSC / Class Code
    6515
    Issuing Office
    FA4417 1 SOCONS
    Primary Contact
    MELISSA PEREZ HUGHES
    State
    FL
    ZIP Code
    32544-5810
    AI Product/Service
    product

    Description

    1.0 Product Overview: This document outlines the requirements for a small, lightweight, and durable multi-therapy infusion pump designed for use in pre-hospital, aeromedical, and military environments. The Office of the Command Surgeon is seeking a portable device that offers a wide range of therapeutic applications in a compact and rugged form factor, capable of reliable function in demanding field conditions.  The intended use for this device is for intravascular, subcutaneous, intra-arterial, epidural, and perineural infusions administered by licensed military healthcare professionals in clinical, ambulatory, and medical transport settings (both air and ground).

    2.0 Core Capabilities and Specifications: The primary requirements for the infusion pump are focused on size, weight, and cube (SWaP) to ensure it can be seamlessly integrated into already existing medical pack-outs and ruck sacks.

    Feature                                                                             Required Specification

    • Dimensions                                                            Objective: Less than 6” x 4” x 2”
    • Weight                                                                        Less than 15 oz (excluding battery)
    • Delivery Modes                                                     Must support: Continuous, Patient-Controlled       Analgesia (PCA), Intermittent, Multi-Step, Total     Parenteral (TPN),  and Epidural infusions.
    • Flow Rate                                                                  0.1 mL/h to 999 mL/h
    • Flow Rate Accuracy                                           ±2.5% under normal conditions
    • Volume to be Infused                                       0.1 mL to 9,999 mL
    • Drug Library                                                            Must support up to 40 clinical care areas and at    least 1,000 drugs per area.
    • Power Source                                                         AC: 100-240V, 50-60 Hz DC: Rechargeable Li-Ion battery
    • Battery Life                                                              Minimum of 24 hours of continuous operation at  a rate of 125 mL/h.
      • Environmental                                      Device must feature a touch screen that is fully     operable when wet. The casing must be   designed to withstand drops and fluid splashes.
      • Operating Temperature                    +5°C to +40°C (41°F to 104°F)
      • The device must meet the following minimum standards for military and aeromedical use:
    • Electromagnetic Compatibility: Must meet MIL-STD-461G for Electromagnetic Interference (EMI) and Compatibility (EMC).
    • Aeromedical Certification: Must be certified as "Safe-to-Fly" for use in military and commercial air transport.
    • FDA Clearance: The device must have current FDA clearance for the indicated uses.

    • Technical Considerations: The device must be capable of wired data connectivity. While a non-standard serial port is acceptable, it must be supplied with an appropriate adapter to ensure connectivity with modern computer systems (e.g., USB).
    • Consumables: Please specify if the infusion pump system requires proprietary consumables (e.g., administration sets).
    • Performance Testing: The device will be subject to a series of performance checks measuring the Volume to be Infused (VTBI) to validate accuracy and reliability.
    • Submission: In response to this request for information, vendors shall submit a capability statement or white paper demonstrating that the proposed items meet all outlined specifications. Submissions must be received by the date specified in the SAM.gov announcement and submitted to the point of contact identified therein.

    Key dates

    1. April 8, 2026Posted Date
    2. April 14, 2026Proposals / Responses Due

    AI search tags

    Frequently asked questions

    MULTI-THERAPY INFUSION PUMP is a federal acquisition solicitation issued by DEPT OF DEFENSE. Review the full description, attachments, and submission requirements on SamSearch before the response deadline.

    SamSearch Platform

    Stop searching. Start winning.

    AI-powered intelligence for the right opportunities, the right leads, and the right time.