Closed Solicitation · DEPARTMENT OF VETERANS AFFAIRS

    NORTH TEXAS VAHCS CENTRIFUGE REPLACEMENT

    DEPARTMENT OF VETERANS AFFAIRS
    Sol. 36C25726Q0181SolicitationSet-aside: Total Small Business Set-Aside (FAR 19.5)San Antonio, TX
    Closed
    STATUS
    Closed
    closed Mar 6, 2026
    POSTED
    Feb 10, 2026
    Publication date
    NAICS CODE
    339112
    Primary industry classification
    PSC CODE
    6640
    Product & service classification

    AI Summary

    The Central Texas VA Health Care System is seeking Service-Disabled Veteran Owned Small Business sources for the purchase of two cytology centrifuges. The equipment must meet specific operational and safety requirements and be delivered within 120 days of contract award. Interested vendors must respond by March 6, 2026.

    Contract details

    Solicitation No.
    36C25726Q0181
    Notice Type
    Solicitation
    Set-Aside
    Total Small Business Set-Aside (FAR 19.5)
    Posted Date
    February 10, 2026
    Response Deadline
    March 6, 2026
    NAICS Code
    339112AI guide
    PSC / Class Code
    6640
    Primary Contact
    Joshua Weisman
    State
    TX
    ZIP Code
    78240
    AI Product/Service
    product

    Description

    The Central Texas VA Health Care System (CTXVAHCS), Temple Hematology Department in Pathology & Laboratory Medicine Service is seeking Service-Disabled Veteran Owned Small Business (SDVOSB), Veteran Owned Small Business (VOSB), or Small Business sources capable of meeting the requirement listed below.  The acquisition will be accomplished using commercial item procedures in accordance with FAR Part 12.

    NOTE: TO QUALIFY SDVOSB AND VOSB VENDORS MUST BE REGISTERED AS SUCH IN THE U.S. SMALL BUSINESS ADMINISTRATION DATABASE.

    The North American Industry Classification System (NAICS) is 339112.  Any SDVOSB, VOSB, or Small Business firms who wish to identify their interests and capability to provide this product must provide product specifications, performance and delivery information by notifying the Contract Specialist no later than 1:00 PM Eastern, March 6, 2026. Packages shall be e-mailed to Joshua Weisman via email at joshua.weisman@va.gov.

    Any vendor who responds to this Notice must provide credentials to perform the requirement as outlined in the below scope.

    Disclaimer:

    This Sources Sought Notice is issued solely for information and planning purposes only and does not constitute a solicitation.  All information received in response to this Notice that is marked as proprietary will be handled accordingly.  In accordance with the Federal Acquisition Regulation, responses to this notice are not offers and cannot be accepted by the Government to form a binding contract.  Responders are solely responsible for all expenses associated with responding to this Notice.

    Purpose:

    Requesting the purchase of the following equipment for Temple Hematology Department in Pathology & Laboratory Medicine Services. This is a one-time purchase, and it does not require any additional manufacturer warranty. Purchase new instruments for the Hematology department in Pathology & Laboratory Medicine service to replace End of Lifecycle/Non-Economical repair instruments.

    Scope:

    Two Cytology Centrifuges are due to replacement due to end of lifecycle and non-economical repair index. The replacement instrumentation must be able to act as a centrifuge specifically for cytological specimens and slides. The instrument must operate at 110v with programable cycle times. The rotor speed must be between 100 and 2000 rpm. The instrument must include a locking lid for safety while the rotor is spinning while accommodating cytological slides. The cytological rotor should be included with the centrifuge and include slide brackets. The instrument must fit within a 2ft3 (24”x24”x24”) area with the lid closed and open.

    Location:

    The equipment will be used in the Olin E. Teague Veterans' Center, 1901 Veterans Memorial Drive, Temple, TX 76504-7451

    Salient Characteristics: See table below.

    Common Nomenclature:

    Cytological Centrifuge

    Electrical Data:

    115V AC 10 amperes, 60Hz

    Dimensions, size, capacity:

    24” W x 24” D x 24”H

    Principles of operation:

    Centrifuge and Stainer

    Restrictive environmental conditions:

    20-25 degrees C, relative humidity range of 30-60%

    Intended use:

    Application of centrifugal force to evenly apply staining chemicals onto cytological slides to highlight the necessary homological specimens.

    Other pertinent information that describes the item, material or service required:

    Completely Enclosed Rotor Chamber

    See-Through High-Strength Lid

    Dual Safety Interlock System - To Ensure the Lid Closed and Locked

    Audible End-Of-Run Indicator Is User Settable

    Lock Settings to Prevent Accidental Changes

    Run Timer (1-99 mins)

    Speed Settable in RPM or RCF

    Simultaneous Digital RPM and RCF Display

    Rotor and slide holder adaptors included

    Period of Performance:

    Purchase and delivery of instruments to Olin E. Teague Veterans Center; Pathology & Laboratory Medicine Service; Hematology Department; within 120 days of award of the contract to purchase. No base or option years required.

    Deliverables/Supplies:

    There are no supplies or deliverables with this contract request, equipment only.

    Vendor Security Contract Requirements:

    This vendor requires no access to VANTHCS or any VA information system.

    NO VA DATA OF ANY TYPE SHALL BE TRANSFERRED FROM THE VA.

    PERSONNEL QUALIFICATIONS:

    The Contractor shall be responsible for employing technically qualified personnel to perform the work specified in this SOW. The Contractor shall maintain the personnel, organization, and administrative control necessary to ensure that the work delivered meets the contract specifications and requirements. The history of each contractor employee must contain experience directly related to the task and functions he/she intends to perform under this contract. The Government reserves the right, during the life of the resulting contract, to request work histories on any contractor employee for the purpose of verifying compliance with the above requirements; additionally, the Government reserves the right to review resumes of contractor personnel proposed to be assigned. Personnel assigned to, or utilized by, the Contractor in performance of work shall be fully capable of performing the requirements contained in the /SOWPWS in an efficient, reliable, and professional manner. The normal manner of dress is business casual.

    Specific Personnel Qualifications Requirements:

    Personnel assigned to or utilized by the contractor in the performance of this contract shall, as a minimum: Be qualified by the contractor as capable to perform as a field service representative.

    PERFORMANCE STANDARDS AND QUALITY MEASUREMENT:

    Performance Standards:

    The Government performs surveillance to determine if the Contractor exceeds, meets, or does not meet these standards.  The Government shall use these standards to determine Contractor performance and shall compare Contractor performance to the Acceptable Quality Level (AQL).  The Quality Assurance Surveillance Plan (QASP) method of surveillance will be by random inspection.

    Performance Evaluation:

    After acceptance of the contract the system will be monitored for equipment performance including but not limited to; safety (recall Notifications and changes in reagents and equipment notifications), quality assurance (emergency repairs response, hot line troubleshooting response, quality of results compared to peer groups and CAP performance) and documentation (invoice accountability and invoice verification, equipment specification and user manuals), customer satisfaction (response to complaints and laboratory employees training).

    Operating Constraints:

    The instrument must be operational with 110v with a programable rotor speed of up to 2000rpm; it must be able to accommodate at least eight cytological slides for spinning, have a function to program run times, and the lid must lock for safety. The instrument must fit within a 2ft3 (24”x24”x24”) area with the lid closed and open.

    DISCLAIMER: THIS NOTICE OF INTENT IS NOT A REQUEST FOR COMPETITIVE QUOTES.

    However, Contractors may submit a capability statement, proposal, or quotation which shall be considered by the contracting officer whether to continue with the sole source procurement or to provide for competition. Due to technical issues, please allow additional time if you’re using electronic means.

    Responses received will be evaluated; however, a determination by the Government to award proposed procurement based upon responses received to this notice is solely within the discretion of the Government.

    Key dates

    1. February 10, 2026Posted Date
    2. March 6, 2026Proposals / Responses Due

    AI search tags

    Frequently asked questions

    NORTH TEXAS VAHCS CENTRIFUGE REPLACEMENT is a federal acquisition solicitation issued by DEPARTMENT OF VETERANS AFFAIRS. Review the full description, attachments, and submission requirements on SamSearch before the response deadline.

    SamSearch Platform

    Stop searching. Start winning.

    AI-powered intelligence for the right opportunities, the right leads, and the right time.